Recently, a diagnostic panty-liner kit—「AL-SENSE」®(Commonsense, Caesarea, Israel) was developed to de- tect amniotic fluid leak. Its mechanism of action is de- scribed elsewhere [1]. The panty-liner has several ad- vantages. Firstly, it is worn by the women for some hours, so it reflects a relatively lengthy time period and not a single point in time, which increases its sensitivity in detecting minor amniotic fluid leakage. In addition, it contains a unique polymer that is capable of differenti- ating amniotic fluid from urine.
In previous third trimester clinical studies [1,2], the AL-SENSE panty-liner test was documented as a sensi- tive tool for diagnosing premature rupture of the mem- branes (PROM) and differentiating amniotic fluid leaks from urinary incontinence. One study showed 100% test, microscopic evaluation for ferning and ultrasound determination of intra-uterine amniotic volume [2]. These negative and 87.1% positive predictive values, with a sensitivity of 100% and specificity of 75%. Another study showed 94.7% negative and 87% positive predic- tive value with a sensitivity of 96% and specificity of 84%. The panty-liner was compared to a series of con- ventional tests such as speculum examination, nitrazine results were confirmed by another study [3] comparing the panty-liner test to speculum examination for detec- tion of ruptured membranes; it found 98% sensitivity and 65% specificity.
Those studies enrolled women mostly at the third tri- mester. However, since amniotic fluid leakage during the second trimester has even more serious complica- tions and poor outcomes than during the third trimester, the present study was undertaken to establish the AL-SENSE reliability in the second trimester and to validate the stability time of the positive color of AL-SENSE stains even after exposure to very small of early second trimester amniotic fluid. The amniotic fluid was drawn at genetic amniocentesis between 18 and 23 weeks of gestation.
The study was approved by the local Institutional Re- view Board (Helsinki committee) and written informed consent was obtained from each patient prior to enrollment.
The AL-SENSE comprises a regular panty-liner and is worn as any other hygienic panty-liner. Inside the panty-liner the AL-SENSE contains one strip of (color) indicator that is sensitive to pH levels. The indicator strip reacts immediately to any contact with amniotic fluid or other vaginal discharge of pH 5.2 or greater. Observing a change in the color of the pH indicator in- dicates either a leak of amniotic fluid or a urine leak. The indicators were placed in a drying box and observed to determine the stability of the stain.
During amniocentesis performed for medical reasons (primarily age over 35 years), between 18 and 23 weeks of gestation, 0.5 ml of amniotic fluid, of the 30 ml usu- ally drawn, were used for this study.
Two volumes of amniotic fluid, 100 µl and 400 µl, were dripped onto two AL-SENSE panty liners num- bered 1 and 2, respectively. After five minutes the indi- cator strip was removed from the AL-SENSE and the strips were placed in the drying units.
The color status was recorded from both panty-liners immediately, after 5 and 10 minutes, and then every 10 minutes for the first 30 minutes and within 1, 2, 6 and 12 hours after removing the indicator strip and placing it in the drying unit.
3.Statistical Analysis
Statistical analyses were performed using SAS® v9.1 (SAS Institute, Cary N.C., USA). Descriptive statistics are presented in a table. Study results are presented as counts or percentages together with a 95% exact bino- mial confidence interval when relevant.
4.Results
Fifty (50) women were enrolled in the study; 49 (98%) completed it. One patient was excluded due to a protocol deviation.
The distribution of gestational age is presented in Ta- ble 1; most women were in their 18th -19th week of ges- tation.
One hundred percent of samples in both volumes (95% CI: (92.75% - 100%)) changed color from yellow to blue within five minutes and the color remained stable
Table 1. Distribution of gestational age byweeks.| GestationalWeek | N | % | | 16 | 1 | 2.0 | | 17 | 3 | 6.1 | | 18 | 16 | 32.7 | | 19 | 21 | 42.9 | | 20 | 5 | 10.2 | | 21 | 1 | 2.0 | | 22 | 1 | 2.0 | | 23 | 1 | 2.0 | | Total | 49 | 100.0 |
|
|
throughout the twelve hour follow-up period. Initial in- dicator color change did not occur at time 0 for all the samples. In the 100 μl subgroup 7 of the 49 samples, or 14.3% (95% CI: (5.94% - 27.24%)), had a delayed color change and in the 400 μl subgroup 1 of 49, i.e. 2.04% (95% CI: (0.05% - 10.85%)), showed a delay in color change.
5.Discussion
In the present study the AL-Sense panty-liner was found to be very sensitive and changed its color in response to a very small amount of second trimester amniotic fluid obtained at amniocentesis. The color change was stable for 12 hours, the time determined in advance as the completion of the study. So far, the validity of the AL-Sense panty liner has been proven in clinical studies in diagnosing premature rupture of membranes in the third trimester [1-3]. However, the effectiveness of this test had not been tested in second trimester pregnancies.
This study utilized amniotic fluid obtained during amniocentesis. Therefore, the effectiveness of this test should be verified in a clinical setting, as the color of the panty-liner may change in reaction to vaginal secretions when infections that cause elevated pH are present (as stated by the manufacturer). Nevertheless, Mulhair et al. [3] found that women with positive high vaginal swabs did not differ significantly and nine women with positive high vaginal swab had a negative test (no color change).
The high pH level of amniotic fluid (7.42) even at very early stages of pregnancy [4], the ability of the panty-liner to detect it, even in very small amounts (100µl), and the long term stability of the stain makes it a very useful tool in diagnosing cases with premature rupture of the membranes in the second trimester. This period of pregnancy may be associated with significant management dilemmas due to the high percentage of fetal, neonatal and maternal complications with PROM at that time [5-7]. The high sensitivity of this test in clinical studies (100%) makes it very helpful in reassur- ing patients that membranes have not ruptured.
A drawback of this test is its low specificity; hence, when the test is positive the diagnosis should be verified by other means. However, in this study, as we dripped amniotic fluid directly on the panty-liner and there was no contact with vaginal discharge or other fluide, we did not include a positive/negative control.
Another possible use of this test is in patients experi- encing post-amniocentesis fluid leakage. Since most of these cases may be followed expectantly [8,9] the test can be used to indicate cessation of amniotic fluid leak- age (when a positive test become negative).
In conclusion, the panty-liner is effective in the sec- ond trimester of pregnancy and can be used at that time to detect small amounts of amniotic fluid leakage, and to exclude its presence if the leakage consists of urine. It is very sensitive and detects small amounts of amniotic fluid drawn during amniocentesis.
As the present study has been conducted in vitro, the efficacy of AL-Sense should be further verified in clini- cal studies of women with fluid leakage during the sec- ond trimester, in the same way as it has been tested in women during the third trimester.
6.Conflict of Interests
Hadar Kesari-Shoham is employed by the manufacturer of the AL-SENSE panty-liner.
REFERENCES
[1] J. Bornstein, A. Geva, I. Solt, V. Fait, A. Schoenfeld, H.
K. Shoham and J. Sobel, 「Nonintrusive Diagnosis of Premature Ruptured Amniotic Membranes Using a Novel Polymer,」 American Journal of Perinatology, Vol. 23, No. 6, 2006, pp. 351-354. doi:10.1055/s-2006-947159
[2] J. Bornstein, G. ohel, Y. Sorokin, K. Z. Reape, O. Shmaider, H. Kessary-Shoham and E. Ophir, 「Effective- ness of a Novel Home-Based Testing Device for the De- tection of Rupture of Membranes,」 American Journal of Perinatology, Vol. 26, No. 1, 2009, pp. 45-50. doi:10.1055/s-0028-1095183
[3] L. Mulhair, J. Carter, L. Poston, P. Seed and A. Briley, 「Prospective Cohort Study Investigating the Reliability of the AmnioSenseTM Method for Detection Of Spontaneous Rupture of Membranes,」 An International Journal of Ob- stetrics and Gynaecology, Vol. 116, No. 2, 2009, pp. 313-318. doi:10.1111/j.1471-0528.2008.01828.x
[4] E. Jauniaux, D. Jurkovic, B. Gulbis, W. P. Collins, J. Zaidi and S. Campbell, 「Investigation of Acid-Base Bal- ance of Coelomic and Amniotic Fluids in Early Human Pregnancy,」 American Journal of Obstetrics and Gyne- cology, Vol. 170, No. 5, 1994, pp. 1365-1369.
[5] A. Greenough and M. Blott, 「Neonatal Outcome after Prolonged Rupture of the Membranes Starting in the Sec- ond Trimester,」 Archives of Disease in Childhood, Vol. 63, No. 10, 1988, pp. 1146-1150.doi:10.1136/adc.63.10_Spec_No.1146
[6] A. Farooqi, P. A. Holmgren, S. Engberg and F. Serenius, 「Survival and 2-Year Outcome with Expectant Manage- ment of Second-Trimester Rupture of Membranes,」 Ob- stetrics and Gynecology, Vol. 92, No. 6, 1998, pp. 895-901. doi:10.1016/S0029-7844(98)00287-7
[7] J. U. Hibbard, M. C. Hibbard, M. Ismail and E. Arendt, 「Pregnancy Outcome after Expectant Management of Premature Rupture of the Membranes in the Second Tri- mester,」 The Journal of Reproductive Medicine, Vol. 38, No. 12, 1993, pp. 945-951.
[8] R. B. Gold, G. L. Goyert, D. B. Schwartz, M. I. Evans and L. A. Seabolt, 「Conservative Management of Second Trimester Post Amniocentesis Fluid Leakage,」 Obstetrics and Gynecology, Vol. 74, No. 5, 1989, pp. 745-747.
[9] A. F. Borgida, A. A. Mills, D. M. Feldman, J. F. Rodis and J. F. X. Egan, 「Outcome of Pregnancies Complicated by Ruptured Membranes after Genetic Amniocentesis,」 American Journal of Obstetrics and Gynecology, Vol. 183, No. 4, 2000, pp. 937-939. doi:10.1067/mob.2000.108872