This kit uses Sandwich-CLIA as the method. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human Bcl-2 . Standards or samples are added to the appropriate micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human Bcl-2 and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well successively and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human Bcl-2, biotinylated detection antibody and Avidin-HRP conjugate will appear fluorescence. The Relative light unit (RLU) value is measured spectrophotometrically by the Chemiluminescence immunoassay analyzer. The RLU value is positively associated with the concentration of Human Bcl-2. The concentration of Human Bcl-2 in the samples can be calculated by comparing the RLU of the samples to the standard curve.
Assay type | Sandwich |
Format | 96T |
Assay time | 4.5h |
Reactivity | Human |
Detection method | Chemiluminescence |
Detection range | 62.50-4000 pg/mL |
Sensitivity | 37.50 pg/mL |
Sample volume | 100uL |
Sample type | Serum, plasma and other biological fluids |
Repeatability | CV < 15% |
This kit recognizes Human Bcl-2in samples. No significant cross-reactivity or interference between Human Bcl-2and analogues was observed.
As the RLU values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration(pg/mL) | RLU | Average | Corrected |
---|---|---|---|
4000 | 5753460628 | 59081 | 59046 |
2000 | 2528930429 | 27859 | 27824 |
1000 | 1428612682 | 13484 | 13449 |
500 | 65146696 | 6605 | 6570 |
250 | 33643122 | 3243 | 3208 |
125 | 16931469 | 1581 | 1546 |
62.50 | 725785 | 755 | 720 |
0 | 3535 | 35 | -- |
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Human Bcl-2were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Human Bcl-2were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 214.89 | 456.63 | 1932.39 | 213.61 | 468.80 | 1808.51 |
Standard deviation | 20.93 | 44.89 | 182.61 | 20.74 | 50.86 | 184.65 |
C V (%) | 9.74 | 9.83 | 9.45 | 9.71 | 10.85 | 10.21 |
The recovery of Human Bcl-2spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
---|---|---|
Serum (n=5) | 86-98 | 92 |
EDTA plasma (n=5) | 96-111 | 104 |
Cell culture media (n=5) | 93-104 | 98 |
Samples were spiked with high concentrations of Human Bcl-2and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) | EDTA plasma (n=5) | Cell culture media (n=5) | ||
---|---|---|---|---|
1:2 | Range (%) | 91-106 | 91-109 | 99-112 |
Average (%) | 97 | 99 | 106 | |
1:4 | Range (%) | 91-107 | 89-99 | 98-112 |
Average (%) | 98 | 94 | 104 | |
1:8 | Range (%) | 89-106 | 95-109 | 96-111 |
Average (%) | 97 | 101 | 103 | |
1:16 | Range (%) | 97-114 | 89-104 | 85-96 |
Average (%) | 104 | 95 | 91 |
An unopened kit can be stored at 4"C for 1 month. If the kit is not used within 1 month, store the items separately according to the following conditions once the kit is received.
Item | Specifications | Storage |
---|---|---|
Micro CLIA Plate(Dismountable) | 8 wells X12 strips | -20"C, 6 months |
Reference Standard | 2 vials | |
Concentrated Biotinylated Detection Ab (100X) | 1 vial, 120 uL | |
Concentrated HRP Conjugate (100X) | 1 vial, 120 uL | -20"C(shading light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 4"C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | |
HRP Conjugate Diluent | 1 vial, 14 mL | |
Concentrated Wash Buffer (25X) | 1 vial, 30 mL | |
Substrate Reagent A | 1 vial, 5 mL | 4"C (shading light) |
Substrate Reagent B | 1 vial, 5 mL | 4"C (shading light) |
Plate Sealer | 5 pieces | |
Product Description | 1 copy | |
Certificate of Analysis | 1 copy |
Reactivity: | Human |
ELISA Type: | CLIA |
SouthernBiotech/Goat Anti-Human Lambda-Alexa Fluor® 555/2070-32/1.0 mg
vectorlabs/Biotinylated Aleuria Aurantia Lectin (AAL)/B-1395/1 mg
vectorlabs/VECTASTAIN® Elite® ABC-HRP Kit (Peroxidase, Mouse IgG)/PK-6102/1 kit
vectorlabs/VECTASTAIN® Elite® ABC-HRP Kit (Peroxidase, Universal)/PK-6200/1 kit
vectorlabs/Unconjugated Musa Paradisiaca (Banana) Lectin (BanLec)/L-1410/5 mg
GiottoBiotech/Calmodulin N60D/5 mg/G02CLM60cn